How Many Hours Are Really In A Day?

We recently blogged on FDA's draft guidance, Chemical Analysis for Biocompatibility Assessment of Medical Devices, which describes chemical characterization methods that may be used to demonstrate biocompatibility of a medical device as an alternative to conducting certain biological testing.

United StatesFood, Drugs, Healthcare, Life Sciences

We recently blogged on FDA's draft guidance, Chemical Analysis for Biocompatibility Assessment of Medical Devices, which describes chemical characterization methods that may be used to demonstrate biocompatibility of a medical device as an alternative to conducting certain biological testing. We noted that the draft guidance's discussion on determining a device's contact duration raised concern as it contradicted ISO 10993-1:2018, Biological evaluation of medical devices —Part 1: Evaluation and testing within a risk management process. By way of background, FDA's general guidance for biocompatibility, Use of International Standard ISO 10993-1, "Biological evaluation of medical devices – Part 1: Evaluation and testing within a risk management process", references ISO 10993-1 for identification of a device's nature and duration of contact, including cumulative effects with repeated use. The ISO 10993-1: 2018 standard, in turn, describes contact duration as the