On January 31, 2023, the U.S. Food and Drug Administration (FDA)
published a Warning Letter[1] to RightEye, LLC (RightEye), the manufacturer
of the RightEye Vision System, for misbranding and adulteration.
The RightEye Vision System is a Class II Nystagmograph medical
device which is cleared under its 510(k) notification for the
following indication: "recording, viewing, and analyzing eye
movements in support of identifying visual tracking impairment in
human subjects."2
The violations were observed during an FDA inspection on June
28-July 8, 2022 on the Form FDA 483. RightEye responded on July 29,
2022, and FDA concluded the responses were inadequate. In the
Warning Letter, FDA wrote that, for the reasons discussed below,
the device was: (i) misbranded due to promotion beyond its cleared
indications for use, (ii) misbranded due to a failure or refusal to
furnish required material and information, and (iii) adulterated
due to a failure to conform with medical device